Recall details
| Product | Centrella Smart+ Bed, Catalog Number P7900 |
|---|---|
| Brand / manufacturer | Hill-Rom, Inc. |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-2246-2020 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 30 April 2020 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: The Bed Exit System may fail to send a remote alert through the nurse call system if a remote alert was previously sent and cancelled at the in-room nurse call wall unit.
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.