[Recalled] CADD-Solis HPCA Ambulatory Infusion Pump; Model/Number: 21-2111-0402-09-KYO 21-2111-0402-09-KYO CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP 21-2111-0300-09_LOAN CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FO
Authorities recorded a safety problem with this product. The details below come directly from the government recall notice.
Recall details
| Product | CADD-Solis HPCA Ambulatory Infusion Pump; Model/Number: 21-2111-0402-09-KYO 21-2111-0402-09-KYO CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP 21-2111-0300-09_LOAN CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR CMSA 21-2111-0300-09_CMSA CADD Solis PIB PUMP FOR CLINICAL EVALUATION |
|---|---|
| Brand / manufacturer | Smiths Medical ASD, Inc. |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-1667-2025 |
| Country of notice | United States |
| Risk level | Class I |
| Hazard type | recall |
| Notice date | 10 April 2025 |
| EAN / barcode | 10610586034787 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: Affected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.
This Class I recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Stop using the product immediately and follow the authority's remedy (return, repair, or disposal).
Read the full official notice →
Related recalls from Smiths Medical ASD, Inc.
This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.