[Recalled] BBL Chocolate II Agar- IVD for the isolation and cultivation of Neisseria species. Catalog (Ref) Number: 221267 — Boxes contain a reduced concentration of Nicotinamide Adenine Dinucleotide and… | FDA 2022-11-21
This is an official product-safety recall record. Below are the verified details from the issuing authority, in plain language.
Recall details
| Product | BBL Chocolate II Agar- IVD for the isolation and cultivation of Neisseria species. Catalog (Ref) Number: 221267 |
|---|---|
| Brand / manufacturer | Becton Dickinson & Co. |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-0473-2023 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 21 November 2022 |
| EAN / barcode | 20221125 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: Boxes contain a reduced concentration of Nicotinamide Adenine Dinucleotide and does not provide enough of the blood factors necessary to support Haemophilus growth. The reduction could cause a false negative result for Haemophilus isolation in a clinical specimen.
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →
Related recalls from Becton Dickinson & Co.
This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.