[Recalled] Atellica CH 930 Analyzer, Siemens Material Number 11067000, Software Versions V1.19.2 or 1.20.0 The Atellica CH 930 Analyzer is an automated, clinical chemistry analyzer designed to perform in vitro diagnostic tests on clinic

recall fda device united-states

Authorities recorded a safety problem with this product. The details below come directly from the government recall notice.

Recall details

ProductAtellica CH 930 Analyzer, Siemens Material Number 11067000, Software Versions V1.19.2 or 1.20.0 The Atellica CH 930 Analyzer is an automated, clinical chemistry analyzer designed to perform in vitro diagnostic tests on clinical specimens.
Brand / manufacturerSiemens Healthcare Diagnostics, Inc.
Categorydevice
Issuing authorityFDA_DEVICE
ReferenceZ-0862-2020
Country of noticeUnited States
Risk levelClass II
Hazard typerecall
Notice date11 October 2019
EAN / barcode00630414002163

Why it was recalled

According to the FDA_DEVICE notice, the identified hazard is: On the Atellica CH 930 Analyzer, when scanning barcodes at the module console for Integrated Multisensor Technology (IMT) system fluids (Standard A (Std A) and Standard B (Std B)) for Sodium (Na), Potassium (K), and Chloride (Cl), the system will translate the decimal point separators to commas. This will result in invalid concentration values for the fluids, therefore the system will utilize default nominal values instead of lot specific IMT fluid concentration values for calculation of the calibrations and results for the Na, K, and Cl methods.

This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.

What to do if you own this product

Check whether your unit is affected and follow the authority's guidance before further use.

Read the full official notice →

Related recalls from Siemens Healthcare Diagnostics, Inc.

This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.

Shop Recalled Atellica 930 Analyzer
Search on Amazon ›Find on eBay ›
As an Amazon Associate and eBay Partner, we may earn from qualifying purchases.

Explore by topic

Looking up a product? Use ProductGuru — free barcode & product database with 35M+ products. Search by EAN / UPC barcode.
Was this page helpful?