[Recalled] Atellica CH 930 Analyzer, Siemens Material Number 11067000, Software Versions V1.19.2 or 1.20.0 The Atellica CH 930 Analyzer is an automated, clinical chemistry analyzer designed to perform in vitro diagnostic tests on clinic
Authorities recorded a safety problem with this product. The details below come directly from the government recall notice.
Recall details
| Product | Atellica CH 930 Analyzer, Siemens Material Number 11067000, Software Versions V1.19.2 or 1.20.0 The Atellica CH 930 Analyzer is an automated, clinical chemistry analyzer designed to perform in vitro diagnostic tests on clinical specimens. |
|---|---|
| Brand / manufacturer | Siemens Healthcare Diagnostics, Inc. |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-0862-2020 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 11 October 2019 |
| EAN / barcode | 00630414002163 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: On the Atellica CH 930 Analyzer, when scanning barcodes at the module console for Integrated Multisensor Technology (IMT) system fluids (Standard A (Std A) and Standard B (Std B)) for Sodium (Na), Potassium (K), and Chloride (Cl), the system will translate the decimal point separators to commas. This will result in invalid concentration values for the fluids, therefore the system will utilize default nominal values instead of lot specific IMT fluid concentration values for calculation of the calibrations and results for the Na, K, and Cl methods.
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →
Related recalls from Siemens Healthcare Diagnostics, Inc.
This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.