[Recalled] Allura Xper F020 DR Table, System Code 722023 Vascular, cardiovascular and neurovascular imaging applications, Cardiac imaging applications including diagnostics, Non-vascular interventions — There is a potential for the Moni
This is an official product-safety recall record. Below are the verified details from the issuing authority, in plain language.
Recall details
| Product | Allura Xper F020 DR Table, System Code 722023 Vascular, cardiovascular and neurovascular imaging applications, Cardiac imaging applications including diagnostics, Non-vascular interventions |
|---|---|
| Brand / manufacturer | Philips Electronics North America Corporation |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-3232-2018 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 16 July 2018 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.