Recall details
| Product | Aesculap S4 Element 4.5mm Polyaxial Screws (ST240T, ST241T, ST242T, ST243T, ST244T, and ST245T). The S4 Spinal System is intended for anterior/anterolateral and posterior, non-cervical pedicle and non-pedicle fixation. Fixation is limited to skeletally mature patients and is intended to be used a |
|---|---|
| Brand / manufacturer | Aesculap Implant Systems LLC |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-2333-2012 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 25 July 2012 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: There is a possibility that the screw channel that holds the rod in place may be out of tolerance. When the depth of the channel is not deep enough, the rod may not be able to provide ample compression on the insert preventing the screw head from locking down to the bone screw.
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →
Related recalls from Aesculap Implant Systems LLC
This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.