[Recalled] ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP), Material Number 11319151, in vitro diagnostic use in the quantitative determination of total protein in human urine and cerebrospinal fluid — There is a potential for A
A safety authority has issued a recall or warning for this item. The confirmed facts from the source notice are laid out below.
Recall details
| Product | ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP), Material Number 11319151, in vitro diagnostic use in the quantitative determination of total protein in human urine and cerebrospinal fluid |
|---|---|
| Brand / manufacturer | Siemens Healthcare Diagnostics, Inc. |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-2451-2023 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 25 July 2023 |
| EAN / barcode | 00630414279176 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: There is a potential for ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein reagent carryover impacting Enzymatic Creatinine_2 (ECRE_2). Falsely depressed ECRE_2 results may be observed when the assay is processed after the UCFP test on ADVIA Chemistry systems.
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →