[Recalled] Access 2 Reaction Vessels, individual, disposable, plastic tubes used to process patient, quality control and calibration samples on the Access 2 Immunoassay system, Catalog Number 81901 — It has been determined that certain lots
The product on this page is subject to an official recall. We summarise the regulator's findings and the recommended action.
Recall details
| Product | Access 2 Reaction Vessels, individual, disposable, plastic tubes used to process patient, quality control and calibration samples on the Access 2 Immunoassay system, Catalog Number 81901 |
|---|---|
| Brand / manufacturer | Beckman Coulter, Inc. |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-0889-2026 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 07 November 2025 |
| EAN / barcode | 15099590180249 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: It has been determined that certain lots of Access 2 Reaction Vessels may contain manufacturing deformities. The bottom (round end) of the Reaction Vessel may have a small protrusion or hair like deformity. This could cause causing instrument errors (e.g., pipetting, rake, wash carousel errors) that could delay patient results.
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.