[Recalled] 20mm addEASE(TM) Binary Connector with 17 Ga. Needle for use with PAB(R) Containers, BC2000 Product is intended for fluid transfer between a Partial Additive Bag (PAB(R)) container and a drug vial. — Possible solution leakage
A safety authority has issued a recall or warning for this item. The confirmed facts from the source notice are laid out below.
Recall details
| Product | 20mm addEASE(TM) Binary Connector with 17 Ga. Needle for use with PAB(R) Containers, BC2000 Product is intended for fluid transfer between a Partial Additive Bag (PAB(R)) container and a drug vial. |
|---|---|
| Brand / manufacturer | B. Braun Medical, Inc. |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-0424-2014 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 04 April 2013 |
| EAN / barcode | 0061275133 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: Possible solution leakage at the interface of the addEASE(TM) Binary Connector (Product Codes BC2000 and BC1300) and the PAB(R) bag. Based on extensive root cause investigation, it has been determined that the leakage is the result of a misaligned cannula (needle) on the addEASE(TM) Binary Connector, which pierces the PAB(R) bag stopper to access the solution. The misaligned cannula may result i
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →