[Recalled] 2008 K2 Hemodialysis Machine with software version 5.40, Models: (1) Hemodialysis SYS OLC/Diasafe PLS 190610, UDI 00840861100859 (+Serial Number), (2) Machine Short Cap, OLC.DP HP 190630, UDI 00840861100866 (+Serial Number) nd
This is an official product-safety recall record. Below are the verified details from the issuing authority, in plain language.
Recall details
| Product | 2008 K2 Hemodialysis Machine with software version 5.40, Models: (1) Hemodialysis SYS OLC/Diasafe PLS 190610, UDI 00840861100859 (+Serial Number), (2) Machine Short Cap, OLC.DP HP 190630, UDI 00840861100866 (+Serial Number) ndicated for acute and chronic dialysis therapy. |
|---|---|
| Brand / manufacturer | Fresenius Medical Care Renal Therapies Group, LLC |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-1278-2018 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 14 December 2017 |
| EAN / barcode | 00840861100866 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: When the recirculation ultrafiltration (UF) Goal is set to a value greater than 200ml in service mode and the user starts a treatment using the SLED (Sustained Low Efficiency Dialysis) program the display will show an invalid message and the UF pump will run at the recirculation UF rate which may be up to 4000ml/hour. The SLED program is required to limit the UF rate to a maximum of 1000 ml/hour.
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →
Related recalls from Fresenius Medical Care Renal Therapies Group, LLC
This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.