[Recalled] 00597102010 Provisional CR ART SURF PROV 12/PURPLE 10, Rx, Sterile; 00597102012 Provisional CR ART SURF PROV 12/PURPLE 12, Rx, Sterile; 00597102014 Provisional CR ART SURF PROV 12/PURPLE 14, Rx, Sterile; 00597102017 P
This is an official product-safety recall record. Below are the verified details from the issuing authority, in plain language.
Recall details
| Product | 00597102010 Provisional CR ART SURF PROV 12/PURPLE 10, Rx, Sterile; 00597102012 Provisional CR ART SURF PROV 12/PURPLE 12, Rx, Sterile; 00597102014 Provisional CR ART SURF PROV 12/PURPLE 14, Rx, Sterile; 00597102017 Provisional CR ART SURF PROV 12/PURPLE 17, Rx, Sterile; 00597102020 |
|---|---|
| Brand / manufacturer | Zimmer, Inc. |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-2042-2013 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 07 June 2013 |
| EAN / barcode | 00597102010 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →
Related recalls from Zimmer, Inc.
This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.