Product Wiki · Recalls · Health Canada — Medical Devices

StatStrip® Glucose And Beta-Ketone Hospital Meter — recall 82503

Health Canada — Medical Devices · Canada · 2026-08-19 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
82503
Date of alert
2026-08-19
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Chemistry
Product type
Medical devices
Reported risk: A software issue in StatStrip (Gen 2) Glucose and Beta-Ketone Hospital Meters may retain previous patient information during barcode scanning, potentially assigning results to the wrong patient. Users should verify patient identification before accepting scanned data and follow the correction instructions provided by Nova Biomedical.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/statstripr-glucose-and-beta-ke

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.