Product Wiki · Recalls · Health Canada — Medical Devices

MED-RX Wound Care Kit — recall 82219

Health Canada — Medical Devices · Canada · 2026-06-18 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked on this page. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
82219
Date of alert
2026-06-18
Country / jurisdiction
Canada
Authority classification
Type II
Notice type
recall
Product category
General hospital and personal use
Product type
Medical devices
Reported risk: CHS was notified by BD Canada that they are conducting a recall of specific lots of the BD Chloraprep™ Clear - 1 ml and Clear 1.5 ml applicators. The BD Chloraprep™ Clear - 1 ml and Clear 1.5 ml applicators with paper lidding are being recalled due to potential breach of sterility in the packaging, due to wrinkles in the paper lidding which may extend to the seal area. CHS Ltd uses these Chloraprep applicators in the medical devices they manufacture.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/med-rx-wound-care-kit

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.