Product Wiki · Recalls · Health Canada — Medical Devices

Olympus MAJ-1443 and MAJ-1444 Valves — recall 81697

Health Canada — Medical Devices · Canada · 2026-03-05 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
81697
Date of alert
2026-03-05
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Gastroenterology and urology - General hospital and personal use
Product type
Medical devices
Reported risk: Olympus is informing that the MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OERPro and OER-Elite automated endoscope reprocessors. This means the MAJ-1443 and MAJ-1444 can no longer be high-level disinfected using the OER-Pro or the OER-Elite and must instead be reprocessed by methods described in updated IFU.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/olympus-maj-1443-and-maj-1444-

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.