Product Wiki · Recalls · Health Canada — Medical Devices

Automated Impella Controller — recall 81652

Health Canada — Medical Devices · Canada · 2026-02-25 · Type I

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
81652
Date of alert
2026-02-25
Country / jurisdiction
Canada
Risk level
Type I
Risk type
recall
Product category
Cardiovascular
Product type
Medical devices
Reported risk: Abiomed has identified that Automated Impella Controller (AICS) with software V8.2.3, when used in conjunction with an Impella 5.5 with smart assist set (product code 0550-0004), has a delayed "purge system blocked" alarm. Software versions post-v10.1 have this issue addressed. Abiomed recommends that you do not use the AIC with an Impella 5.5 with smart assist set (product code 0550-0004) unless AIC software is updated to v10.1 or above.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/automated-impella-controller

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.