Product Wiki · Recalls · Health Canada — Medical Devices

Hamilton-C2 Intensive Care Ventilator — recall 81503

Health Canada — Medical Devices · Canada · 2026-01-19 · Type I

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
81503
Date of alert
2026-01-19
Country / jurisdiction
Canada
Risk level
Type I
Risk type
recall
Product category
Anaesthesiology
Product type
Medical devices
Reported risk: Hamilton Medical AG became aware of a malfunction with HAMILTON-C2 ventilators when used long term with neonatal patients. If an affected HAMILTON-C2 ventilator is used cumulatively for 91 days with neonatal patient without a restart of the device, the device will switch to "ambient state", resulting in an immediate stop of ventilation. During the "ambient state" the device is alarming audibly and visually until it is switched off by the user. Hamilton Medical AG previously communicated this issue through FSCAs with reference CER 110972 (May 2023) and FSCA-2025-01-01 (January 2025). This updated FSN is provided with a new reference number and addresses only HAMILTON-C2 where the neonatal option is installed. Since the initiation of the first field action in May 2023, Hamilton Medical AG did not receive any new complaint related to the issue for HAMILTON-C2 devices.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/hamilton-c2-intensive-care-ven

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.