A.T.S.® 5000TS (Tourniquet System) With Hoses And Lop Sensor… — recall 81483
Health Canada — Medical Devices · Canada · 2026-01-14 · Type II
As described by Health Canada — Medical Devices: A.T.S.® 5000TS (Tourniquet System) With Hoses And Lop Sensor (Personalized Pressure)
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: Zimmer Biomet is conducting a phased serial specific medical device recall for A.T.S. 5000TS Tourniquet System. Multiple complaints have been received related to the devices in scope indicating that when the device is used in certain international languages (including French), the user interface can become corrupted due to insufficient ram when loading a third font set. This may cause the screen to freeze or stop responding to touch, requiring the device to be restarted. The affected devices can continue to be used in English language mode which is unaffected. The defectiveness was discovered on October 28, 2025. Issue was identified through the complaints handling process. There have been 3 reported complaints in Canada regarding this issue.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.