Health Canada — Medical Devices · Canada · 2025-11-26 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: Beckman Coulter has confirmed that specific diluent lot numbers of Coulter DxH Diluent contribute to elevated platelet (PLT) Daily Checks background counts cycle when used with the UniCel DxH 600/690T/800/900 analyzers. This may result in elevated PLT background counts in Daily Checks ranging from 3.0 to 9.1 x 10 3 cells/μL.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.