Health Canada — Medical Devices · Canada · 2025-11-12 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: It has been found that if there is slow leakage of refrigerant gas from the superconducting magnet over a long period of time, air ingress into the helium vessel via the leak point may result in ice forming from moisture in the air. In the worst case, if the ice is not eliminated, the helium tank could rupture, resulting in a helium leak.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.