Product Wiki · Recalls · Health Canada — Medical Devices

CARESCAPE ONE and Portrait VSM — recall 78359

Health Canada — Medical Devices · Canada · 2025-10-23 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
78359
Date of alert
2025-10-23
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Cardiovascular
Product type
Medical devices
Reported risk: Potential loss of monitoring on CARESCAPE B450, CARESCAPE ONE, B1x5M/B1x5P VSP3.0, B1x5M/B1x5P VSP4.0 patient monitors and portrait vital signs monitor when powered by certain batteries.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/carescape-one-and-portrait-vsm

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.