Product Wiki · Recalls · Health Canada — Medical Devices

Monnal™ T60 Ventilator — recall 78347

Health Canada — Medical Devices · Canada · 2025-10-22 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
78347
Date of alert
2025-10-22
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Anaesthesiology
Product type
Medical devices
Reported risk: The Monnal EVA V2 (2nd generation - expiratory valve) used on Monnal T60/T60 Advanced ventilators can lead to under-delivery of volumes in "child/infant" patients due to poor membrane sealing. No patient incidents have been reported; the detection occurred during tests. Air liquide medical systems is recalling these valves to replace them with the first generation to prevent the risk of hypoventilation.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/monnaltm-t60-ventilator

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.