Product Wiki · Recalls · Health Canada — Medical Devices

Allura and Azurion Systems — recall 77918

Health Canada — Medical Devices · Canada · 2025-08-15 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
77918
Date of alert
2025-08-15
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Radiology
Product type
Medical devices
Reported risk: The 1 Phase UPS is an optional component that can be installed in Allura and Azurion systems to support a controlled shutdown of the PCs (the computers that control the system) in the event of a mains power failure. The 1 Phase UPS provides temporary power to the PCs, allowing the system to save the data of the last acquired run before the PCs shut down. Philips has become aware of a potential safety issue with Philips Azurion and Allura systems configured with a 1 Phase uninterruptible power supply (UPS). Philips has identified that some internal components of the 1 Phase UPS may fail. When this occurs, there will be a complete loss of power to the system, causing the system to shut down and/or not start up. The 1 Phase UPS system may fail without warning and prior to any mains power failure. If the system does not start up or the system shuts down, there is a potential risk of delay or termination of the procedure. The potential delay in treatment and/or termination of the procedure may result in serious adverse health outcomes, including the possibility of death, especially when the system is used with critical patients, particularly those undergoing complex and/or urgent interventions for potentially life-threatening conditions (e.g., acute ischemic stroke, st-segment elevation myocardial ischemia). Recall start date: July 31, 2025
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/allura-and-azurion-systems

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.