Product Wiki · Recalls · Health Canada — Medical Devices

Merit basixCOMPAK™, PhD™ Hemostasis Valve, and basixSKY™ Inflation… — recall 77904

Health Canada — Medical Devices · Canada · 2025-08-14 · Type II

As described by Health Canada — Medical Devices: Merit basixCOMPAK™, PhD™ Hemostasis Valve, and basixSKY™ Inflation Syringe

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
77904
Date of alert
2025-08-14
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Cardiovascular - General hospital and personal use
Product type
Medical devices
Reported risk: Merit Medical became aware of a manufacturing defect after an increased number of complaints. Investigation into the issue concluded that the lack of fluorinated silicone in the bleed back seal (the secondary valve) caused by the mix of incorrect silicone during production line influenced the performance of the PhD™ Hemostasis Valve. The lack of silicone was found to prevent the secondary valve from releasing and closing properly. The valve cap potentially would remain depressed causing leakage. Recall start date: July 30, 2025
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/merit-basixcompaktm-phdtm-hemo

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.