Reported risk: Edwards has confirmed a small number of occurrences impacting the OptiSite Arterial Perfusion Cannula in which a 3mm to 4mm section of wire, from the wire-reinforcement coil at the cannula tip, was found to be released from the cannula body. No customer complaints were reported in Canada. As certain femoral arterial cannula models are made of the same components as the impacted OptiSite Arterial Perfusion Cannula, the scope of this notice includes these femoral arterial cannula models. Recall start date: June 3, 2025
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.