MicroMyst applicator and extended tip applicators — recall 77582
Health Canada — Medical Devices · Canada · 2025-06-11 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: The decision to conduct a recall of the MicroMyst applicator and extended tip applicators was based on internal investigation that demonstrated an incomplete bioburden assessment and an incomplete sterilization location transfer documentation to ensure the effectiveness of the sterilization process. The investigation also identified that the extended tip applicators, exclusively, could potentially contain out-of-specification levels of endotoxin. Recall start date: June 2, 2025
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.