Product Wiki · Recalls · Health Canada — Medical Devices

CombiStar, SafeStar and TwinStar Filters/ HMEs — recall 77579

Health Canada — Medical Devices · Canada · 2025-06-10 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
77579
Date of alert
2025-06-10
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Anaesthesiology
Product type
Medical devices
Reported risk: Users have reported a slow increase of the CO 2 -curve when the gas sample line of the gas measurement of a Dräger anesthesia machine is connected to the gas sampling port of the affected filters/HME. The comparatively slower increase in the CO 2 -curve only occurs when the gas sample line is connected to the gas sampling port of an affected filter/HME. Other setups are not affected. Recall start date: May 23, 2025
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/combistar-safestar-and-twinsta

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.