Newport HT70 Ventilator Plus — recall 77491

Health Canada — Medical Devices · Canada · 2025-05-22 · Type I

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
77491
Date of alert
2025-05-22
Country / jurisdiction
Canada
Risk level
Type I
Risk type
recall
Product category
Anaesthesiology
Product type
Medical devices
Reported risk: The C92 capacitor of HT70 ventilators may fail, disabling the output of the power management chip causing the ventilator to cease operation and stop ventilation. An audible alarm will be initiated, and the safety valve will open. In addition to the C92 issue, HT70 ventilators containing C159 capacitors reported failures of the shutdown buzzer alarm test, and as a result, the alarm "hardly sounded" or was of insufficient volume or duration per specification, placing a patient at risk of respiratory failure due to unexpected loss of ventilation. Recall start date: May 21, 2025
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/newport-ht70-ventilator-plus

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.