Beckman Coulter Au Immunochemistry System - Apolipoprotein B — recall 77411
Health Canada — Medical Devices · Canada · 2025-05-07 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: The Apo B lot 2631 VAS LOSR6143_2631_ML.01 contains an incorrect target value of 166 mg/dl for the level 4 calibrator ODR3022, lot 3000. The correct level 4 target value is 116 mg/dl. The value in g/l unit is correct. Recall Start Date: April 21, 2025
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.