Comprehensive® Shoulder System Primary Mini Length Shouldler Stem — recall 77112
Health Canada — Medical Devices · Canada · 2025-03-13 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: Zimmer Biomet is conducting a lot specific medical device recall for the Comprehensive® Shoulder Primary Stem (7mm Micro & 7 mm Mini). Five complaints reported that the stem implant did not fit in the inserter or that the inserter could not be detached from the stem. There were no injuries reported associated with the five complaints received. The internal investigation identified excess metal on the inner taper of the stem due to a chipped manufacturing tool. Recall start date: March 10, 2025
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.