In2Flow Nasal Cannula, With OD15 Adapter — recall 77025
Health Canada — Medical Devices · Canada · 2025-02-26 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: Hamilton Medical AG received several complaints where the plug (the one at the opposite side of the tubing) became detached from the In2Flow Nasal Cannula, causing leakage due to unintentional disconnection of that part. This may result in degraded high flow support and oxygen delivery, which could lead to a certain level of patient o2 desaturation. Affected product: Hamilton Medical In2Flow Nasal Cannulas with a production date dated after 31.12.2022 and till January 2025 of all sizes and all lot numbers. Recall start date: February 10, 2025
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.