Product Wiki · Recalls · Health Canada — Medical Devices

MassCheck® and 3PLUS1® Antimycotic Drugs/Extended Plasma Controls — recall 76984

Health Canada — Medical Devices · Canada · 2025-02-20 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
76984
Date of alert
2025-02-20
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Chemistry
Product type
Medical devices
Reported risk: New data from stability tests can not confirm Chromsystems instruments & Chemicals GmbH previous information regarding the storage life of calibrator 92051/XT and controls 0253/XT, 0254/XT after reconstitution at storage temperatures of +20 to +25 °c and +2 to +8 °c provided in the instructions for use and on the package inserts. In addition, adsorption effects may occur if the reconstituted products are stored in glass containers. Recall start date: February 5, 2025
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/masscheckr-and-3plus1r-antimyc

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.