Product Wiki · Recalls · Health Canada — Medical Devices

Stryker Precision Thin Blade Attachments — recall 76837

Health Canada — Medical Devices · Canada · 2025-01-16 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
76837
Date of alert
2025-01-16
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Orthopaedics
Product type
Medical devices
Reported risk: During a review of the complaint handling and risk assessment data (October, 2024) it was identified that the precision thin blade attachments have a potential to be out of measurement specifications, preventing them from securely fitting into the compatible handpiece. Recall start date: January 8, 2025
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/stryker-precision-thin-blade-a

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.