Health Canada — Medical Devices · Canada · 2024-12-24 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: For certain Siemens XA images, the isocenter calibration value is calculated incorrectly due to an incorrect value for distance source to isocenter that is saved by Siemens in the standard dicom attribute (0018,1111). The correct value is stored in a Siemens private dicom attribute (0021, xx17, "Siemens sms-ax acq 1.0") in the dicom data. Incorrect calibration can lead to errors in the measurements provided by qAngio XA 3D. Recall start date: December 16, 2024
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.