Health Canada — Medical Devices · Canada · 2024-12-11 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: The Zeus IE features two gas measurement modules, a patient gas analyzer, and a system gas analyzer. Both contain a pump which is sampling the gas to be analyzed. This sampling pump has shown an increased failure rate, which might lead to different effects during use of the Zeus IE. The effects depend on the therapy settings. Recall Start Date: July 31, 2024
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.