Product Wiki · Recalls · Health Canada — Medical Devices

High-Speed Burrs — recall 76566

Health Canada — Medical Devices · Canada · 2024-11-22 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
76566
Date of alert
2024-11-22
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
General and plastic surgery
Product type
Medical devices
Reported risk: Holes in the inner barrier system have been detected by the laboratory during a visual inspection for the product quality. Due to the compromised sterile packaging the affected product is being recalled. Recall Start Date: November 11, 2024
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/high-speed-burrs

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.