Product Wiki · Recalls · Health Canada — Medical Devices

BD BBL Sensi Disc Ampicillin 2 Μg (AM-2) — recall 76548

Health Canada — Medical Devices · Canada · 2024-11-20 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
76548
Date of alert
2024-11-20
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Microbiology
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: BD identified through potency test, as part of special stability test requests to monitor Ampicillin AM-2, unsatisfactory results of 65% at 18 months. The potency acceptance criteria is 80-140% and product shelf life is 24 months. A retest with control batch and extended sampling confirmed the low potency results. Recall Start Date: November 6, 2024
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/bd-bbl-sensi-disc-ampicillin-2

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.