Product Wiki · Recalls · Health Canada — Medical Devices

Plum Infusion Systems — recall 76494

Health Canada — Medical Devices · Canada · 2024-11-13 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
76494
Date of alert
2024-11-13
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
General hospital and personal use - General and plastic surgery
Product type
Medical devices
Reported risk: ICU Medical has received reports of counterfeit CSB batteries being used with Plum infusion systems. While these counterfeit batteries are visually similar to the Plum batteries supplied by ICU Medical, they are in fact not the same batteries and have not been tested or validated for use with Plum infusion systems. Preliminary reports suggest that these counterfeit batteries fail to hold their charge and the pump may display messages to replace batteries earlier than expected. Do not use spare parts that are not authorized by ICU Medical. Please ensure that any entity providing service or repair activities for your plum infusion systems uses only parts authorized or directly provided by ICU Medical. Recall start date: October 21, 2024
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/plum-infusion-systems-0

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.