Giraffe™ OmniBed Carestation CS1 — recall 76201

Health Canada — Medical Devices · Canada · 2024-09-25 · Type II

As described by Health Canada — Medical Devices: Giraffe™ OmniBed Carestation CS1

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
76201
Date of alert
2024-09-25
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
General hospital and personal use
Product type
Medical devices
Reported risk: GE Healthcare has become aware that for certain Giraffe™ OmniBed and Giraffe™ OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This may lead to the doors becoming loose. If the doors become loose, the system activates a high priority alarm, and the canopy movement will stop. Recall start date: September 17, 2024
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/giraffetm-omnibed-carestation-

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.