Health Canada — Medical Devices · Canada · 2024-08-30 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: Roche received complaints from US customers for Creatine Kinase (CK) lot 755471 (expired June-2024) on cobas c 311/501/502 and INTEGRA 400 plus about abnormal calibration signals, flagged calibrations (Sens.E) and control recovery issues. Subsequent investigation identified a positive bias at the lower measuring range and a negative bias at the higher measuring range. The issue affects a small number of cassettes of the mentioned lot. Recall Start Date: August 19, 2024
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.