BD Facsduet™ Sample Preparation Systems — recall 75970
Health Canada — Medical Devices · Canada · 2024-08-27 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: Through a customer complaint investigation, BD identified a specimen volume bias issue with the BD Facsduet™ sample preparation system when using software version 1.4.1. When the multidispense feature is enabled in the BD Facsduet™ sample preparation system software, and the user is preparing a two-tube assay, the specimen volume dispensed into the first tube will be at least 3% higher than the specimen volume dispensed into the second tube. For assays, where absolute counts are calculated using BD Trucount™ tubes, an increase in the volume may result in higher-than-expected absolute counts in the first tube compared to the second tube. This issue will be observed when preparing the following assays on the BD Facsduet™ sample preparation system: 1. BD Multitest™ IMK kit with BD Trucount™ tubes. note: if using the BD Multitest™ IMK kit with BD Trucount™ tubes, a QC message will appear on your report if the bias is outside the acceptable range per assay requirements. 2. BD Multitest™ 6-color TBNK with BD Trucount™ tubes only if consecutive replicates of two are run. 3. for all other quantitative assays utilizing BD Trucount™ tubes in consecutive replicates of two, please follow your institution's procedures to determine if a review of reported results is necessary. This issue does not affect assays that are determining percent positive, semi quantitative information. Recall start date: August 12, 2024
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.