Product Wiki · Recalls · Health Canada — Medical Devices

Azurion 7 M20 — recall 75967

Health Canada — Medical Devices · Canada · 2024-08-26 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
75967
Date of alert
2024-08-26
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Radiology
Product type
Medical devices
Reported risk: Philips has identified that grease leaked from the bearings and/or anti-corrosion oil applied on the bearings can result in excessive lubrication on the ceiling rail. This can reduce the friction between the friction wheel and the rails. The reduced traction force causes the friction wheel to slip on the rail, leading to a "position slip" error mode being detected by the software. When this error occurs, the x-y-rz joystick is unable to send motorized requests to the relevant axes. Due to this issue, the motorized longitudinal movement of the Flexarm stand may be inconsistent (not smooth) and eventually may become unavailable. Recall start date: August 7, 2024
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/azurion-7-m20-1

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.