Health Canada — Medical Devices · Canada · 2025-09-25 · Type II
As described by Health Canada — Medical Devices: Salem Sump™ products
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: Cardinal Health has created an electronic instruction for use (eIFU) for Salem Sump™ products because of reports of breakage in the ARV due to improper use of the device. These changes to the labeling have been made to increase user awareness of how to use the ARV correctly and to inform users of the new warning. Lots manufactured after May 1, 2019 are affected. Recall start date: July 22, 2024 Cardinal Health is issuing an update to the urgent medical device product correction originally issued in July 2024, on all lots of Salem Sump™ product codes listed that were manufactured from August 1, 2020. The purpose of this update is to provide additional guidance on how to properly use the Anti-Reflux Valve (ARV).
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.