Health Canada — Medical Devices · Canada · 2024-08-01 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: Customers have observed an issue where the metal part of the sample holder detaches from the plastic holder base when the robot arm grips a sample tube. If this occurs at a centrifuge, the metal part can inadvertently enter the centrifuge along with the sample tube. The centrifuge is unable to detect the presence of the detached metal part and begins to rotate. This can pose risks to both the equipment and the samples being processed. Recall start date: July 22, 2024
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.