Infinity 5 and 7 Implantable Pulse Generator — recall 75782
Health Canada — Medical Devices · Canada · 2024-07-03 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: Abbott has initiated a voluntary medical device recall of Implantable Pulse Generator (IPG) for patients with non-rechargeable Infinity™ Deep Brain Stimulation (DBS) neurostimulation systems, as the duration between the IPG reaching battery elective replacement indicator (ERI) threshold and end of service (EOS) may be 45-55% shorter than indicated in the product labeling. Recall Start Date: June 6, 2024
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.