Product Wiki · Recalls · Health Canada — Medical Devices

AXIOM Vertix and AXIOM Multix — recall 75669

Health Canada — Medical Devices · Canada · 2024-06-14 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
75669
Date of alert
2024-06-14
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Radiology
Product type
Medical devices
Reported risk: Siemens Healthineers was informed by the monitor supplier that 5 out of 27.000 delivered power supplies of a touch-display power supply unit were affected by overheating due to a short circuit. In one of the five events, the connected monitor caught fire. No injuries were associated with these events. To ensure the continued safe operation of your system and to safeguard its lifetime, the power supply of the touch display will be replaced preventively. Recall start date: May 29, 2024
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/axiom-vertix-and-axiom-multix

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.