Product Wiki · Recalls · Health Canada — Medical Devices

VITROS Immunodiagnostic Products Anti-HBs Calibrators And Reagent — recall 75591

Health Canada — Medical Devices · Canada · 2024-05-30 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
75591
Date of alert
2024-05-30
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Immunology
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: QuidelOrtho's internal stability testing of VITROS anti-HBs calibrators detected a gradual reduction in calibration signal for calibrator levels 2 and 3 over the shelf life of the product for the specified lots. This reduction in signal may affect the shape of the calibration curve generated during calibration, compared to the calibration curve generated earlier in the product's shelf life. A change in the shape of the calibration curve may cause patient and QC samples tested using the associated VITROS® immunodiagnostic products anti-HBs reagent pack to report results as falsely elevated, potentially causing negative results to report as borderline, and borderline results to report as positive. This issue is detectable by performing QC as results generated will be higher than expected and/or outside the expected limits for QC, depending on the age of the lot and amount of signal reduction experienced in calibrator levels 2 and 3. Recall start date: May 21, 2024
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/vitros-immunodiagnostic-produc

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.