7392 PCP Folding Bathtub Transfer Bench — recall 75372
Health Canada — Medical Devices · Canada · 2024-04-10 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: The recall is due to the buttons on certain legs not securing properly. Airway Surgical Appliances Ltd was alerted to this issue by one of their customers, who informed them that a user experienced a collapsed leg while attempting to use the product. Recall start date: March 27, 2024
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.