Digital Radiography Mobile X-Ray System Dx-D 100 — recall 75221
Health Canada — Medical Devices · Canada · 2024-03-12 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: A component supplier of the Agfa Dx-D 100 informed Agfa about a potential safety issue. The issue as described by the supplier is related to the steel cable installed inside the mobile column which supports the weight of the telescopic arm with the tube head. The potential risk consists of an interference which could cause that the safety system which blocks the arm if the cable is cut off does not work properly. The interference occurs between the steel cable and a metal plate of the arm's carriage due to the shape of the cable's end. Recall Start Date: February 23, 2024
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.