Product Wiki · Recalls · Health Canada — Medical Devices

Origio Sperm Wash — recall 75188

Health Canada — Medical Devices · Canada · 2024-03-07 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
75188
Date of alert
2024-03-07
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Obstetrics and gynaecology
Product type
Medical devices
Reported risk: During testing it was found that one of the batches, had an insufficient concentration of sodium pyruvate - only 0.002 mg/ml compared to a specified concentration of 0.046-0.137 mg/ml (1.5-4.3% of the specified amount). Part of this formulation batch had also been used to create three lots of commercially released sperm wash medium. Recall start date: January 22, 2024
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/origio-sperm-wash

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.