Captus® 4000e Thyroid Uptake System, all models — recall 74833
Health Canada — Medical Devices · Canada · 2024-04-19 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: In 2023, three Captus® 4000e systems incurred a spring arm failure, causing the arm to fall. The collimator and arm, together, weigh 45 lbs (20.4 kg), which carries a risk of injury if the collimator comes into contact with the technologist or the patient. Recall start date: November 26, 2023
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.